Micro Thermal Mass Flow Meter for Small Flow

The thermal mass flowmeter is a flow meter based on the principle of thermal diffusion. The device uses the constant temperature difference method to measure the gas accurately. It has the advantages of small size, high digitization, easy installation, accurate measurement and so on.

  • Using MEMS flow sensor chip, the sensor has the characteristics of high precision, high sensitivity and strong anti-interference.
  • This product has display screen and setting buttons, which is easy to operate and read directly.(optional)
  • The zero point stability of the sensor is greatly improved compared with the ordinary thermal flow meters.
  • Full range high stability.
  • Full range precision and excellent repeatability.
  • Combined with structural optimization, the flow meter can greatly reduce pressure loss and energy consumption compared with
  • traditional mechanical instruments.
  • LCD display instantaneous flow and cumulative flow, clear and intuitive,easy to read.
  • 4~20mA standard signal output and second impulse output to choose.
Enquiry

Micro gas mass flow meters is specially designed for the measurement and process control of various types of small flow gases. This series of
sensors are made with advanced micro-electro-mechanical system (MEMS) flow sensing chips, and are suitable for all types of clean gases. The Unique packaging technology makes it suitable for a wide range of pipe sizes, low cost, easy installation, no temperature and pressure compensation, and can replace traditional volumetric or differential pressure flow meters

The sensor part of the thermal gas mass flow meter consists of two reference platinum resistance. When the flow meter is in the working state, a temperature sensor continuously measures the medium temperature T1. The other sensor will be heated to a temperature T2. And this sensor is a speed sensor to check the flow rate. T2 is greater than T1. There is a temperature difference △T=T2-T1.

When gas flow going through, the gas molecules in the process will take away temperature, so that the value of T2 has been decreased. So if you want to make sure △T unchanged, it must increase the speed sensor power supply current. The greater the velocity of the medium, the greater the power supply current of the speed sensor. At the same time, there is a specific functional relationship between the velocity of the gas and the heat, that is the connotation of the constant temperature difference principle.

V=[K(Q/△T)^1.87]/ρg  …………(1)

 

In the formula:ρg — specific gravity of fluid (related to medium density)

V — Flow rate

K – equilibrium coefficient

Q — heat added (related to specific heat)

△T — temperature difference

Since the sensor temperature is always about 30 ° C higher than the medium (ambient) temperature automatically, the thermal gas mass flow meter does not need temperature compensation in principle.

The thermal gas mass flow meter is suitable for the medium temperature range of -40-220℃.

ρ=ρn*[(101.325+P)/101.325]*[(273.15+20)/(273.15+T)]……(2)

 

In the formula:ρ — medium density under working conditions (kg/m3)

ρn — medium density under standard conditions (101.325kPa, 20℃) (kg/m3)

P — Operating pressure (kPa)

T — Operating temperature (℃)

As we can see equation (1) and (2),  the function of medium flow rate and operating pressure, gas density and operating temperature has been determined.

The thermal mass flow meter using the principle of constant temperature difference, so it needn’t temperature and pressure compensation.

Specifications

Precision(%)±(1.5+0.2FS)Medium temperature– 10 ~ 55
Response                         time (MS)50-1000.Humidity<95%RH(no    frost,                   no

icing)

Maximum working pressure (Mpa)1.0ConnectorG3/8, G1/2, G3/4
Working power15v-24v 100mAThe     overall     power consumption< 2.4 W.
Output model4-20mA,               pulse

(optional)

Communication methodsRS485(Modbus

protocol)

DisplayInstantaneous

flow, cumulative flow

CalibrationAir (20℃, 101.325kPa)

Note: The above data were measured at 20℃, 101.32kPa, dry air

Mechanical dimensions (mm)

micro thermal mass flow meter size